21
OPTIMASS 2000 / 2010C with or without heating jacket / insulation
Ambient Temp.
T
amb
°C
Max. medium
temp. T
m
°C
Temperature Class
Max. Surface
Temp. °C
40
65 T6 T80
75 T5 T95
110 T4 T130
130 T1-3 T150
65
75 T5 T95
110 T4 T130
130 T1-3 T150
OPTIMASS 2300C with aluminium converter housing, with or without heating
jacket / insulation
Ambient Temp.
T
amb
°C
Max. medium
temp. T
m
°C
Temperature Class
Max. Surface
Temp. °C
40
50 T6 T80
65 T5 T95
100 T4 T130
130 T3 - T1 T160
50
65 T5 T95
100 T4 - T1 T130
60 60 T4 - T1 T90
65* 65 T4 - T1 T95
OPTIMASS 2300C with stainless steel converter housing, with or without heating jacket /
insulation
Ambient Temp.
T
amb
°C
Max. medium
temp. T
m
°C
Temperature Class
Max. Surface
Temp. °C
40
50 T6 T80
65 T5 T95
100 T4 T130
120 T3 - T1 T150
50
65 T5 T95
75 T4 - T1 T105
55 55 T5 - T1 T85
2.5 OPTIMASS 2000 / 2010C / 2300C
The OPTIMASS 2000 / 2010C / 2300C are suitable for temperature classes T6….T1 according to
the following tables:
* Only for equipment configurations according to the table in section 2.2
The cable supplied by KROHNE is designed for a continuous working temperature of up to
105°C. Alternative cabling must be a heat-resistant type with a continuous working tempera-
ture of 80°C.
AD OPTIMASS 300_010 Haz Areas Rev 01.qxp 05/03/2014 13:24 Page 21