17
(5)Thisinstrumenthasanauto-dischargefunction.
With the test leads connected to the circuit under test,
release the Test button to discharge capacitances in the
circuitaftertest.Confirmthattheindicationonthevoltage
monitorbecomes0V.
#DANGER
●
Donottouchthecircuitundertestimmediatelyaftertesting.
Capacitancesstoredinthecircuitmaycauseelectricshock.
●Leave the test leads connected to the circuit and never
touchthecircuituntiladischargeiscomplete.
Auto-discharge function
This is a function to release the capacitances stored in the
circuit under testautomaticallyafter test.The voltagemonitor
shows the dischargecondition. Thisfunction willbe disabled
bydisconnectingthetestleads2sec.ormorebeforedischarge
iscomplete.
(6)Set the Range switch toOFFposition, and disconnect
thetestleadsfromtheinstrument.