29
Product Service 1-800-522-7658
Chapter 5: Using Your Protector Fume Hoo
d
Work Techniques
Keep all materials at least 6 inches inside of the sash, and
perform all contaminated operations as far to the rear of the
work area as possible.
Segregate all clean and contaminated materials in the work
area.
Avoid using techniques or procedures that disrupt the
airflow patterns of the hood.
Final Purging
Upon completion of work, the hood should be allowed to
operate for two to three minutes undisturbed, to purge
airborne contaminants from the work area before shutting
down blower.
Unloading Materials and Equipment
Objects in contact with contaminated material should be
surface decontaminated before removal from the hood.
All open trays or containers should be covered before being
removed from the hood.
Shutdown
Turn off the fluorescent light and hood blower, then close
the sash.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector XStream 9840400 and is the answer not in the manual?

Labconco Protector XStream 9840400 Specifications

General IconGeneral
BrandLabconco
ModelProtector XStream 9840400
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals