ANALYSIS
1. Press *ION to turn the meter on.
1.3
2. Press *IOK to select Measure. Main Menu
* Measure
Data Logging
Options
16:02:19 01/04/05
3. Rinse a clean tube (0290) three
times with the blank.
For the most accurate results, use
thesametubefortheblankandthe
sample.
4. Fill the tube to the fill line with the
blank. Pour the blank down the
inside of the tube to avoid creating
bubbles.
5. Dry the tube with a lint-free cloth. Put
on a dry positioning ring. Cap the
tube. Wipe the tube thoroughly
again with a lint-free cloth.
ANALYSIS & CALIBRATION 19