SELECTING A TURBIDITY CALIBRATION CURVE
1. Press ON to turn the meter on.
1.3
2. Scroll down and then press *IOK to
select Options.
Main Menu
Measure
Data Logging
* Options
16:02:19 01/04/05
3. Scroll down and then press *IOK
to select Turbidity.
Options
Averaging
* Turbidity
Date/Time
Language
16:02:19 01/04/05
4. Scroll down and then press *IOK
to select Calibration.
Turbidity
Units
* Calibration
16:02:19 01/04/05
5. Scroll down and then press *IOK
to select a Calibration curve. Select
a calibration option based on the
composition of the standards that
will be used to calibrate the meter.
Available options are: Formazin,
Japan Standard, Attenuation.
Turbidity
Units
* Calibration
16:02:19 01/04/05
For the most accurate results, the
Attenuation option should be chosen
when samples are over 500 NTU.
The range for the Attenuation option
is 40–4000 NTU.
Calibration
* Formazin
Japan Standard
Attenuation
16:02:19 01/04/05