TESTING
Getting Started
1. Remove the cap from the bottom of the TRACER to expose the
electrode.
2. Before the first use, rinse the electrode in deionized water and dry.
3. For best results, calibrate for conductivity with a standard in the
expected
range of the sample. For maximum accuracy calibrate
from low conductivity
value standards to high conductivity value
standards.
4. Store dry.
Changing the Measurement Function
The meter can be set to measure Conductivity, TDS (ppm), TDS (mg/L) or
Salinity (ppm). To change the mode:
1. Turn the TRACER on.
2. Press and hold the MODE button for 2 seconds. The display will begin
to scroll through the units.
µS (Conductivity) | ppm (TDS) | mg/l (TDS) | ppm (Salinity “S”)
3. Release the MODE key when the desired mode is displayed.
4. Note that the “HOLD” function can not be on when changing
functions.