Section #4 – Instrument Tests 36
Electronic Copy of LTI's UltraLyte LR B User's Manual - 2
nd
Edition © 2002
Cover screw
Plastic washer
Adjusting screw
1.5 mm
2.5 mm
Scope rear view Scope rear view
Refer to the instructions below to do the Scope Alignment Procedure.
1. Steady the instrument on a steady base.
2. Expose the adjustment screws. Using a 2.5 mm Allen wrench,
remove the cover screws as shown in Figure 25.
Each cover screw assembly includes a small plastic washer.
It is imperative that you not lose the washers. Failing to
replace them renders the scope at risk to water damage.
3. Ensure that the UltraLyte LR B is powered ON.
4. Press the TEST OPTIONS button two times to
activate the Test Tone display screen.
It should look like Figure 24 on page 34.
5. Select a target (see page 35).
6. Scan the target. Press and hold the TRIGGER while panning
the instrument across the target. When the tone achieves its
highest pitch, the laser light beam is hitting the target.
7. Adjust the scope use a 1.5 mm Allen wrench to turn the
adjustment screws. Adjust the position of the scope aiming
dot, as shown in Figure 26.
Figure 25. Scope Adjustment Screws
Figure 26. Adjusting the Scope

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Laser Technology UltraLyte LR B and is the answer not in the manual?

Laser Technology UltraLyte LR B Specifications

General IconGeneral
BrandLaser Technology
ModelUltraLyte LR B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals