Section #8 – Troubleshooting Tips 53
Electronic Copy of LTI's UltraLyte LR B User's Manual - 2
nd
Edition © 2002
Section #8 – Troubleshooting Tips
What You Will See Required Action
Unit powers OFF
by itself.
-or-
No power at all.
Verify that the batteries are installed
correctly.
Replace the batteries.
Keep in mind that the instrument
automatically powers OFF if there is
no instrument activity for a period of
15 minutes.
The in-scope red aiming
dot is not visible.
Press the TRIGGER to activate
the red dot.
Press the BRIGHTNESS CONTROL
button to increase the intensity of the
red dot.
Adjust the polarizing filter.
E01 error code. No
range or speed
readings.
Can you measure to a wall that is
about 50 feet (or 20 meters) away?
Check the scope alignment.
Restore the factory defaults.
When measuring a short range to a
small target, aim slightly above the
target.
Erroneous readings/not
repeatable.
Do you have a clear line of sight?
Check the scope alignment.
Is it raining or foggy. If yes, is the
weather filter active? If not, turn it
ON. Keep in mind that the minimum
range will be 250 feet (76 meters).
When measuring a short range to a
small target, aim slightly above the
target
No serial
communication.
Verify the cable connection.