Unit description
4
3 Unit description
3.1 Unit description
The immersion thermostats E 200 have a device for fixing the immersion thermostat to the bath vessel
(clamping bracket). An adapter is supplied for the deep-drawn LAUDA baths 003, 011, 019 and 025.
The type designation of the Ecoline bath/circulation thermostats consists of the control unit E 200 and the
type of bath.
Example: Control unit E 200 and bath 003 produces Thermostat Type E 203ö.
The letter T (for "Transparent") refers to the baths made of polycarbonate. Type E 203 are supplied with
bath cover. For other baths made of stainless steel bath covers are available as accessory.
( Section 10. Accessories)
3.2 Pumps
All units are supplied with a pressure pump with vario-drive. The pump has an outlet with a rotatable bend
(immersion thermostat) which is connected to the pump nipple for external thermostating circuits
(bath/circulation thermostats). An additional outlet provides circulation within the bath. By turning the
setting knob you can choose between both outlets or divide the flows.
The pump chamber of immersion thermostats is rotatable in a restricted way to reach an optimal
circulation. The pump can be used up to viscosities of 150 mm²/s during heating up. To get an optimum
accuracy of control a viscosity of < 30 mm²/s is recommended.
One of five pump output steps can be selected using the operating menu.
On small bath thermostats (e.g. E 203 or E 206 T) and with operation as bath thermostat it is advisable to
use output step 1. The advantage is a low heat generation while having a uniform circulation, that means,
that the unit can work without cooling down to just above ambient temperature.
When operating as circulation thermostat with an external circuit it is preferable to use a larger flow
setting in order to ensure a small temperature difference, especially at higher temperatures and in
conjunction with oil as the bath liquid.
The pump outflow connection can be closed off without causing any damage to the pump.