5.5 Cooling the thermostats
At bath temperatures down to just above ambient temperature (approx. 2 – 10°C) it is possible to work
without cooling. Additional cooling is required for lower temperatures
Immersion thermostat. ! attach the cooling coil ( Section 5.1.).
Bath/Circulation thermostats ! fitted with cooling coil, as standard..
Cooling possibilities
.
down to 20 °C Mains water ! keep the water consumption as low as possible!
.
down to –30°C flow-through cooler DLK 10/ DLK 25 (depending on bath size and
temperature) Section 10. Accessories
! use water/glycol mixture (ratio 1:1).
Mains supply 230V will be switched off in case of fault. Consumption of
current 2A max.
Use insulated tubing!
When thermostating an external system the equipment must be arranged in the
following order: thermostat ! external circuit ! flow-through cooler ! thermostat