Starting up
34
Turn the potentiometer slowly anticlockwise.
!
The unit
must switch off at the bath temperature.
Step 1 - 2 (see above) must follow.
Set the overtemperature switch-off point again above the
bath temperature and wait until the indication TEMP
appears on the display, then
reset with the key.
6.5.2 Low-level protection and testing
$
$$
$

Double signal tone sounds, if the bath liquid falls so much
that the heater is no longer covered with liquid completely.
1. The display shows LEVEL (low-level) and the fault triangle
is flashing
! heating is switched off on both poles,
! pump is switched off by the electronics
2. Top up the bath Section 5.2 and reset with the key.
If necessary repeat several times in case that the pump does not purge
immediately.
Testing at regular intervals by lowering the bath level. Place a hose on the pump
connector and pump some of the bath liquid into a suitable container.
Step 1 - 2 must follow.
Bath temperature during this test not below 0°C or higher than 50°C, otherwise
danger of burn injuries !
If there is any irregularity when testing the safety devices, switch off the unit
immediately and pull out the mains plug !
Have the unit checked by the LAUDA service or the local service organisation!