Unit description
15.05.01
YAEE0013
7
Protokoll:
The interface operates with 1 stop bit, no parity bit and 8 data bits.
Transfer rate either 2400, 4800, 9600 (factory setting) or 19200 baud as selected.
The RS232 interface can be operated with or without hardware handshake (RTS/CTS).
The command from the computer must be terminated with CR, CRLF or LFCR.
The response of the thermostat is always terminated with CRLF.
CR = carriage return (hex: 0D)
LF = line feed (hex: 0A)
Example::
Transfer of setpoint 30.5°C to the thermostat.
Computer Thermostat
„OUT_SP_00_30.5“CRLF
"
#
„OK“CRLF