5-16 Performance Verification Rev. D
1) Connect the DMM and record the
voltage reading
in Table 6, column
DMM.
2) Press
Clear Sweeps
3) After 10 sweeps, read off the
DSO mean parameter
, and
record the
measurement in Table 5, column
Mean
.
!
For each DC voltage applied to the DSO input, repeat parts 1), 2), 3) and 4).
!
Calculate the
Difference (
)
by subtracting the
DMM voltage
reading from
the
DSO mean
voltage reading. Record the test result in Table 5, and compare
the
Difference (
)
to the corresponding limit in the test record.
!
Repeat step 5.7 Procedure for the other channels, substituting channel controls
and input connector.