Rev. D Performance Verification 5-17
5.8 Offset Accuracy
The offset test is done at 5mV/div, with a signal of
±
1 Volt cancelled by an offset of
the other polarity.
Specifications
±
(0.015 x Voffset + 0.005 x FS + 1mV)
FS [volts] = 8[div] x sensitivity[volts/div]
Voffset [volts] = setting offset voltage
Procedure
!
Recall
xxxP019.PNL
or configure the DSO:
Panel Setups :
Recall FROM DEFAULT SETUP
Channels Trace ON
Channel 1, Channel 2, Channel 3 & Channel 4
Input Coupling :
DC 1M
on all 4 Channels
Input gain : 5
mV/div
on all 4 Channels
Input offset :
+1 Volt
on all 4 Channels
Trigger setup :
Edge
Trigger on :
1
Coupling 1 :
DC
Mode :
Auto
Time base :
2 msec/div
.
Channel use :
4
Record up to :
25 k
Channels Trace OFF
Channel 1, Channel 2, Channel 3 & Channel 4
ANALYSIS CONTROL
Trace ON :
A, B, C & D
Select MATH TOOLS
For Math :
Use at most 5000 points
Redefine A, B, C, D
Channel 1, Channel 2, Channel 3 & Channel 4
Use Math ? :
Yes
Math Type :
Average
Avg. Type :
Summed
For :
100 sweeps
WAVE PILOT :
MEASURE
Mode :
Custom
Statistics :
off
!
Connect the test equipment as shown in Figure 5-4.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy Waverunner2 LT374 Series and is the answer not in the manual?

LeCroy Waverunner2 LT374 Series Specifications

General IconGeneral
BrandLeCroy
ModelWaverunner2 LT374 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals