Hardware User Manual
Envisu C-Series SDOIS / 9054-10011 / Revision N 11
Warning! Optical Radiation
Do not remove the optical fiber while the system is operational. Turn off
power before removing the optical fiber.
Warning! Optical Radiation
Do not operate the system with the fiber optical cable removed. Do not look
directly into the optical fiber port.
Warning! Patient Contact
Do not simultaneously touch the patient and the computer.
Caution! Ventilation
System requires proper ventilation. Do not block the Front or Sides of the
System.
For product warranty by the manufacturer the following must be observed:
• The device must be used according to this manual, “Leica Spectral
Domain Ophthalmic Imaging System. The system should not be
used in any manner aside from that specified in the user manual.
• Installation, assembly, service, and maintenance must be performed
by the manufacturer or an authorized service representative only.
• No modifications to the device are allowed.
• The electrical installation of the room where the device is used
complies with national standards. Do not remove or disable the
ground pin of any power connection.
• Supply cables are for connecting Leica, Inc. equipment only and
may not be used with any other equipment.
Environmental Conditions
The device is intended for operation in the hospital environment. The device may be
used only in enclosed rooms and must be placed on a solid floor.
Warning! Explosion Hazard
Do not operate the device in an environment where explosive and/or
flammable substances are used or stored. In the event of an emergency,
disconnect the power cords immediately.
Caution!
This device is not designed for use in damp or high humidity environments.
Do not allow condensation to form on any component. Do not place fluid-
filled containers on any surface of the device.