EasyManuals Logo

Leica Envisu C Series User Manual

Leica Envisu C Series
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Hardware User Manual
Envisu C-Series SDOIS / 9054-10011 / Revision N 8
probe is 750uW. The system is compliant to IEC60825-1,
Class 1 Laser.
Mode of Operation
Continuous
Power Requirements
North America:
120V, 60 Hz
Europe and Australia:
185-240 Single Phase VAC, 50/60Hz, 400W max. (Not
evaluated by UL)
Japan:
100V, 50 or 60 Hz
Power Cord
North America: Hospital grade NEMA 5-15 cord, 20’.
Europe: CEE 7/7 plug, 3.5m cord.
Australia: AS/NZS 3112:2000 plug 6.1 m cord
Japan: Hospital grade NEMA 5-15 cord, 4.6m
Type of Protection
Against Electrical
Shock
Class I Equipment
Degree of Protection
Against Ingress Of
Water (Foot pedal
only)
Minimum IPX6 Ordinary Equipment
Classification
Class 2 Medical Device (Canada)
Class IIa Medical Device (European Union)
Class IIa Medical Device (Australia)
Class II Medical Device (Japan)
Class 1 Laser Product
Continuous Operation
Not for use in flammable environments
Not provided sterile - Cleaning Instructions provided