Check & Adjust TPS1200+ 99
<No.of Meas:> Shows the number of runs executed. One run consists of
a measurement in face I and face II.
<σ l Comp:> and similar lines show the standard deviations of the deter-
mined adjustment errors. The standard deviations can be calculated from
the second run onwards.
It is recommended to measure at least two runs.
8. MEAS (F5) if more runs have to be added. Continue with step 3.
OR
CONT (F1) to accept the measurements and to proceed to TOOLS
Adjustment Results. No more runs can be added later.