Filter
with
Gas
Vent
Replacement
in
I.V.
Disposable
Sets
Replace
the
filter
with
gas
vent
every
THREE
HOURS
or
when
the
filter
becomes
clogged
or
air
is
slowly
vented.
Replace
with
Level
1
product
number
F-10
or
F-30.
Refer
to
the
disposable
packaging
for
appropriate
filter
replacement
product
number.
|
WARNING
!:
Replacement
filter
with
gas
vent
must
be
fully
primed
before
continuing
infusion.
Thoroughly
check
patient
line
for
air
bubbles
before
opening
roller
clamp.
If
air
‘bubbles
are
present,
disconnect
from
the
patient
and
re-prime
before
continuing
infusion.
ーー
|
1.
Close
all
clamps
on
the
D-series
disposable
set
and
on
new
filter
with
gas
vent.
|
|
2.
Remove
the
used
filter
with
gas
vent
from
the
holder
while
still
connected
to
the
disposable
set.
|
3.
Install
the
new
filter
assembly
into
the
#3
holder.
4.
Using
aseptic
technique,
disconnect
upper
luer-lock
fitting,
and
after
removing
upper
male
luer
end
cap
from
the
new
filter,
connect
the
set
to
new
filter
inlet.
5.
Holding
the
used
filter
horizontally,
disconnect
the
lower
patient
line
luer-
lock.
Connect
the
used
filter
luer-lock
fittings
together
and
discard.
6.
With
the
filter
outlet
tube
clamped,
open
the
clamps
above
the
filter.
The
filter
with
gas
vent
will
self
prime.
7.
Slowly
release
the
outlet
tube
clamp
and
allow
the
outlet
tube
to
fill
completely.
8.
Remove
the
second
luer-lock
end
cap
and
connect
outlet
tube
luer-lock
to
patient
line.
9.
Remove
the
filter
with
gas
its
from
its
holder
and
tap
it
against
the
cabinet
several
times
until
all
trapped
air
bubbles
have
been
vented
out.
Return
the
filter
to
its
holder.
Note:
The
“‘DISPOSABLE”
alarm
will
be
activated
whenever
the
filter
with
gas
vent
assembly
is
removed
from
its
#3
holder.
14