Water
Over
Temperature
Alarm
Press
and
hold
the
Water
Over
Temperature
alarm
test
button
on
the
power
and
test
panel
on
the
front
of
the
unit.
The
display
will
jump
to
45°C
(as
long
as
the
button
is
pressed
on),
and
the
red
over
temperature
indicator
will
illuminate
and
the
audible
alarm
will
start
to
pulse.
NOTE:
The
Water
Over
Temperature
alarm
test
button
fully
tests
the
over
temperature
circuitry
of
the
system,
not
just
the
visible
red
display
indicator
and
the
audible
alarm.
NOTE:
Pressing
the
Water
Over
Temperature
test
button
or
the
Alarm
test
button
will
disable
the
System
1000
even
after
the
buttons
have
been
released.
The
unit
must
be
shut
off
and
then
re-started
to
clear
the
alarm
condition.
Alarm
Testing
The
audible
alarm
and
all
RED
indicator
lights
may
be
tested
by
pressing
the
Alarm
test
button
on
the
power
and
test
panel
on
the
front
of
the
cabinet.
NOTE:
Pressing
the
Alarm
test
button
or
the
Water
Over
Temperature
test
button
will
disable
the
System
1000
even
after
the
buttons
have
been
released.
The
unit
must
be
shut
off
and
then
re-started
to
clear
the
alarm
condition.
Performance
Testing
Periodic
testing
of
the
overall
operating
efficiency
of
the
System
1000
can
be
accomplished
in
one
of
two
simple
ways:
1.
COLD
START
TEST
NOTE:
This
performance
test
requires
a
SIMS
Level
1
non-sterile,
not-for-
patient-use
D-series
disposable
set
to
be
installed
properly
on
the
System
1000.
Store
the
System
1000
fluid
warmer
in
a
room
where
the
room
temperature
is
approximately
21°C
(70°F).
With
a
non-sterile
D-series
disposable
set
in
place,
note
the
time,
and
then
turn
the
power
switch
ON.
The
green
“OPERATING”
indicator
will
illuminate.
Rapidly
rising
numbers
will
appear
on
the
circulating
water
bath
temperature
display.
A
41°C
temperature
reading
will
appear
in
approximately
3
to
7
minutes
on
an
efficiently
operating
unit.
NOTE:
If
the
system
1000
does
not
pass
the
cold
start
test,
it
should
be
returned
to
SIMS
Level
1
or
an
authorized
distributor
of
SIMS
Level
1
for
service.
Do
not
attempt
to
change
the
calibration
settings.
2.
THERMAL
CALIBRATION
WELL
(TCW).
Available
from
SIMS
Level
1,
a
TCW
may
also
be
used
to
test
performance
and
confirm
proper
calibration
of
the
water
bath
temperature.
(Part
#
8003002)
18