Maintenance
92
GA10218_002_C0 - 10/2016 - © Leybold GmbH
by using a vacuum cleaner so that the filter mat can be used again.
In the ventilation line is a dust filter installed. This filter has to be cleaned or changed
when using in dirty environment (cat.-no. 200000683).
4.4.3 Exchanging the Oil
Remove the cover of the mechanical section as described in Chapter 4.4.1.
Electrical hazards
During all maintenance and connection work, make sure that the mains cable have
been reliable disconnected and do not carry a mains voltage.The leak detector
must only be used in with the hoods closed. The electrical connections must only
be provided by a trained electrician as specified, for example, by the regulations EN
50110-1.
When the pump has been pumping hazardous substances, determined the kind of
hazard first and ensure that suitable safety precautions are taken.
Observe all safety regulations
When disposing of waste oil you must observe the applicable regulations for the
safety of the environment.
The oil change procedure are described in the corresponding Operating Instruc-
tions GA 01.601 and these must be followed closely.
As already stated before only Leybonol LVO (Cat. no. L31001) must be used for
the Trivac D 2,5E pump in the PHOENIX L300i.
After completion of all maintenance work the cover of the mechanical section must
engage properly in the openings at the bottom.
4.4.4 Cleaning
The housing of the PHOENIX L300i is made of painted plastic parts. Thus for the
purpose of cleaning, only such agents should be used which are generally also
used for other painted or plastic surfaces (mild household cleaning agents, for
example). Normally a moistened peace of cloth will do. Never use any solvents
which are capable of dissolving paint (like acetone, toluol, etc.).
A soft brush or a vacuum cleaner is recommended for cleaning the ventilation slits.
4.4.5 Exchanging the Fuses
Before exchanging the fuses you must disconnect the mains cord.
1. Switch the PHOENIX L300i off.
2. Pull the mains cord off the PHOENIX L300i.
3. Use a screwdriver to fold out the lid of the mains socket from the right (the mains
switch is not affected by this).
4. The fuses can be removed by pulling the drawers out which are marked by the
arrows. When reinserting these make sure that the arrows point downwards.
5. In any cases two fuses of the same rating must be inserted. The required mains
fuses are: T 10A slow blow (20x5mm dial.) for 100V...230V.
Caution