1.6 Default Settings
The following parameters are set like shown when in the menu of the PHOENIX
L300i under Settings > Parameters, Load / Save is chosen.
Scale linear
Display range: 4 decades
Time axis: 32 seconds
LCD inverted off
Background in standby mode: off
Calibration request: off
Mass: 4 (helium)
Recorder: leak rate
Volume: 2
Leak rate unit: mbar l/s
Mode: Vacuum
Trigger level 1: 1E-9 mbar l/s
Trigger level 2: 1E-8 mbar l/s
Trigger level 3: 1E-7 mbar l/s
Leak rate external test leak (vacuum): 1E-7 mbar l/s
Leak rate external test leak (sniffer): 1E-5 mbar l/s
Vent delay: 2 seconds
Automatic purge (PHOENIX L300i DRY
and PHOENIX L300i MODUL only)
on
Pressure: mbar
Minimum volume: 0
Beep: on
Maximum evacuation time: 30 minutes
Audio Alarm Type: Trigger Alarm
Max. pressure limit for sniff mode 0.15 mbar
Min. pressure limit for sniff mode 0.05 mbar
Control location local
Alarm delay: 30 seconds
Leak rate filter: auto
Zero: enable
Vacuum ranges normal
Upper display limit 1E-5 mbar l/s
Service message exhaust oil filter
(PHOENIX L300i only)
on
Contamination protection Off
Switch-off limit for contamination protection:
(Limits: 1E-6 mbar*l/s … 1E+3 mbar*l/s)
1E-3 mbar*l/s