EasyManua.ls Logo

LEYBOLD PHOENIX L300i - Important Safety Information

LEYBOLD PHOENIX L300i
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety Information
GA10218_002_C0 - 10/2016 - © Leybold GmbH 5
Important Safety Information
Indicates procedures that must be strictly observed to prevent hazards to persons.
Indicates procedures that must be strictly observed to prevent damage to, or des-
truction of the product.
Emphasizes additional application information and other useful information provi-
ded within these Operating Instructions.
The Leybold PHOENIX L300i leak detector has been designed for safe and
efficient operation when used properly and in accordance with these Operating
Instructions. It is the responsibility of the user to carefully read and strictly
observe all safety precautions described in this section and throughout the
Operating Instructions. The PHOENIX L300i must only be operated in the pro
-
per condition and under the conditions described in the Operating Instructions.
It must be operated and maintained by trained personnel only. Consult local,
state, and national agencies regarding specific requirements and regulations.
Address any further safety, operation and/or maintenance questions to our nea
-
rest office.
Failure to observe the following precautions could result in serious
personal injury!
Electrical hazards
During all maintenance and connection work, make sure that the mains cable have
been reliable disconnected and do not carry a mains voltage.The leak detector
must only be used in with the hoods closed. The electrical connections must only
be provided by a trained electrician as specified, for example, by the regulations EN
50110-1.
Avoid exposing any part of the human body to the vacuum. Only handle the leak
detector when the pump is vented.
After a mains power failure the leak detector can run up automatically once more.
Before changing oil or fuses, make sure that the mains cable have been discon-
nected.
The leak detector is not suited for operation in explosion hazard areas.