Programming
248083310-1 Rev 3 4-5
5. Type: <m1> -- the unit’s menu is displayed to show
commands to use to change parameters.
<m1>
*** User Menu ***
<?> Display Parameters
<V> Verbose Mode
<sa> Start Counting
<sb> Stop Counting
<gv> Get Version Number
<ra#> Set Range 1 (counts)
<rb#> Set Range 2 (counts)
<ta#> Set Sample Seconds
<aa#> Set Alarm Channel #
<ab#> Set Alarm Threshold (# of counts)
<ae#> Set Alarm Suppression #
6. The instrument waits 20 seconds after the last command is
received, validated and performed, then begins sampling
ASCII
Programming
Syntax
The REMOTE 5012PN, 5102PN instrument can be programmed by an
ASCII terminal (i.e. Hyperterminal) using the ASCII protocol. The
command format is a start character followed by a command followed
by a terminating character. Any characters before the start character,
or after the terminating character are ignored. White spaces between
command character and command parameter are ignored.
Command Structure
The commands are defined as single case-sensitive ASCII characters.
Format is defined as:
Note: The < and >
characters are part of the
command and must be
typed. The brackets [ and ]
are field delimiters and are
not typed.
<x [yyyy]>
where:
< = Start Character
Note: All command
characters are lower case,
except for the V (verbose).
x = Command Character
yyyy = Optional Command Parameter
> = Terminating Character