Programming
248083201-1 Rev 6 4-7
Range2= 1000.
Service High = 3000 mV
Service Low = 145 mV
Alarm Channel = 1
Alarm Threshold = 1000 counts
Suppress Alarms = 2
Channel 1 = 0.3
Channel 2 = 0.5
5. Type: <m1> -- the unit’s menu is displayed to show commands to
use to change parameters.
<m1>
*** User Menu ***
<?> Display Parameters
<V> Verbose Mode
<sa> Start Counting
<sb> Stop Counting
<gv> Get Version Number
<ra#> Set Range 1 (counts)
<rb#> Set Range 2 (counts)
<ta#> Set Sample Seconds
<aa#> Set Alarm Channel #
<ab#> Set Alarm Threshold (# of counts)
<ae#> Set Alarm Suppression #
6. The instrument will wait 20 seconds after the last command is sent,
then begin sampling
ASCII
Programming
Syntax
Using a simple ASCII protocol, the REMOTE 3012 family of
instruments can be programmed from an ASCII terminal (i.e. Hyper
Terminal). The protocol format is based on a start character, followed
by a command, which is then followed by a terminating character. Any
characters before the start character, or after the terminating character
are ignored. White spaces between command character and command
parameter are ignored.
Command Structure
The commands are defined as single case-sensitive ASCII characters.