Lighthouse REMOTE 2012, 3012, 5012, 3102, 5102 Operating Manual
4-12 248083201-1 Rev 6
SET SAMPLE TIME (seconds)
Note: Sample times <5
seconds are not
recommended.
<ta#>
Sets sample time in # seconds, where # >1 and nnnn < 3600.
Type:
<ta30>
Response:
<ta30>Sample Time = 30 secs
SET ALARM CHANNEL
<aa#>
Sets the alarm channel to channel 1 or 2.
Type:
<aa1>
Response:
<aa1>Alarm Channel= 1
SET ALARM THRESHOLD
<ab#>
Sets Alarm High count threshold for the configured alarm channel.
Value # must be less than range in order to set the threshold. Set the
value to 0 (zero) to disable alarm.
If the count for the configured alarm channel exceeds the threshold
then:
• The red SERVICE indicator on the instrument blinks.
• The EXTERNAL ALARM relay, described below, is set.
• The alarm alert will be seen on the ASCII terminal as illustrated
next:
Type:
<ab1000>