Lighthouse REMOTE 3104P, 5104P Operating Manual
A-4 248083336-1 Rev 3
40034 unsigned integer Sample Time [low]
40035 unsigned integer Data Set [high]. Works in conjunction with 40036. Data entered
here is applied to the device through the command register.
40036 unsigned integer Data Set [low]
40037 unsigned integer Alarm Mode. Type of alarming performed
40038 unsigned integer Alarm Parameter. Control parameter for given alarm mode.
40039 unsigned integer Laser Reference Voltage (millivolts)
40040 unsigned integer View Volume. Divide by 100 to get percentage.
For example: 6550d = 65.50%
40041 ASCII string Flow Unit. Defines unit as cfm, lpm, mlpm char[0], char[1]
(NULL terminated string).
40042 ASCII string Flow Unit. char[2], char[3]
40043 unsigned integer Calibration Reference Voltage (millivolts)
...
40049 signed integer Printer Options
40050 signed integer Device Options: Bit 3 - Number of Locations (1-400 locations
with four character names, 0-200 locations with eight character
names)
...
40199 unsigned integer Number of Available location names (0=not supported).
Registers 40200-40999: If number of available locations is 200, Registers 40200-40999 used for
8 character names associated with location index.
If number of available locations is 400, Registers 40200-40999 used for 4 character names
200 Locations Available (Register 40050 Bit 3=0)
40200 ASCII string Location_1_char[0], char[1] (NULL terminated string)
40201 ASCII string Location_1_char[2], char[3]
40202 ASCII string Location_1_char[4], char[5]
40203 ASCII string Location_1_char[6], char[7]
...
40996 ASCII string Location_200_char[0], char[1] (NULL terminated string)
Table A-3 Sensor Settings Registers
Register Data Type Description