MODBUS Register Map v1.48, Rx104P
248083336-1 Rev 3 A-5
Alarm Mode (40037) defines the type of calculation performed to
define an alarm condition. Alarm Mode = 0 corresponds to
conventional threshold alarming; channel bit set if threshold exceeded
for that given channel.
Alarm Parameter (40038) defines additional parameters that may be
needed in defining an alarm mode.
Printer Options (40049)
displays the configuration of the instrument’s
printer function.
If Bit-1 of Register 40049 is set, the instrument will print the last
recorded data at the end of each sample. This feature cannot be enabled
if the One Second Data Update feature is enabled.
40997 ASCII string Location_200_char[2], char[3]
40998 ASCII string Location_200_char[4], char[5]
40999 ASCII string Location_200_char[6], char[7]
400 Locations Available (Register 40050 Bit 3=1)
40200 ASCII string Location_1_char[0], char[1] (NULL terminated string)
40201 ASCII string Location_1_char[2], char[3]
...
40998 ASCII string Location_400_char[0], char[1]
40999 ASCII string Location_400_char[2], char[3]
Table A-3 Sensor Settings Registers
Register Data Type Description
Table A-4 Printer Options
Bit Description
0 Unused - non-writable
1 Print on Sample (1=Enabled, 0=Disabled)
2-15 Reserved