10.3 Night Bed Lightning
It is possible to equip Image 3 B with undercarriage lightning. The lightning helps the patient or hospital personnel
to better orientate in room with lowered or turned off light. The lowered intensity of lightning is set up after turning
the bed on. For additional informations about bed lightning see chapter “Bed Illumination”.
10.4 Bed Illumination
The bed is equipped with three-phase illumination:
1. Full intensity of illumination
2. Lowered intensity of illumination
3. Illumination is turned off
The lowered intensity of illumination is set up after the bed is turned on.
After pressing the GO button:
The handset and bed illumination will light up at full intensity.
After elapsing the GO period:
The illumination intensity of handset will be lowered.
7 minutes after GO period has elapsed:
The intensity of bed illumination will be lowered.
In the event any button is pressed (supervisor, handset or side rail) outside of GO period:
The bed illumination will light up at full intensity for 10 minutes and after that, the illumination will be
lowered.
Simultaneously, the handset will light up at full intensity for 7 seconds. After that the illumination will
be lowered.
Turning off all illumination:
Press button (19) and all illumination on the bed (handset and bed illumination) will be turned off.
This function is not blocked by the activation GO button.