What to do if the mattress is not inflating on my LINET Multicare LE?
- AAdam SchultzAug 20, 2025
If the mattress of your LINET Medical Equipment is not inflating, ensure the SCU mains switch is on (the green mains power switch should illuminate). Check the bedframe's mains connection and that the mains plug is correctly inserted. Also, inspect the mattress pipe connection for air leaks, ensure the CPR valves are closed, and check the air pipes for blockages or kinks. If the mattress is only partially inflated, make sure it's unfolded and flat. If the problem persists, it may be due to a faulty power source, SCU, or a damaged mattress, in which case you should notify the manufacturer's service department.