73
24.3.6 Castors
Clean the castors completely.
Grease the castors if necessary (Caro EP 2 by DEA or an equivalent grease)
Check that the castors work properly:
Forward Movement
Unrestricted Movement
Braked
Have the brakes adjusted if necessary.
Have any defective castors replaced.
24.3.7 Accessories
Check that all accessories (for example, lifting pole, siderails, infusion stand, etc.) are working properly.
Replace if necessary.
24.4 Safety Checks
In accordance with §6 of the Medical Devices Operator Ordinance, the operator is required to perform a technical
safety check on the hospital bed every 12 months.
The procedure for performing the safety check is stipulated in VDE 0751 or IEC 601.1.
The manufacturer will give a certificate to service organisations in which the manufacturer declares that the ser-
vice organisation is qualified to perform maintenance on LINET
®
products.
LINET
®
makes every reasonable effort within research, design and manufacturing of beds to ensure that their
products are fit for purpose and product to the highest level of quality. However LINET
®
can take no responsibility
for any damage caused to the product or harm to patients, staff or other individuals as a result of:
Not following the instructions for use, including warning and caution statements, provided in their user manu-
als,
Use of the product for other than its intended purpose as stated in any documentation provided by LINET
®
NOTE On request, the manufacturer will provide circuit diagrams, component part lists, descriptions, calibra-
tion instructions etc. for service personnel for the repair of ME equipment designated by the manufacturer as
repairably by service personnel.
Risk of injury due to incorrect safety checks!
Ensure that safety checks are performed by seller´s customer service or
authorised per- sonnel (certified by the manufacturer) only.
Ensure that the safety checks are recorded in the service and maintenance log.
Warning
Risk of injury due to defective bed!
Have a defective bed repaired immediately.
If the defect cannot be repaired, do not continue to use the bed.
Warning

Table of Contents

Other manuals for LINET Multicare LE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare LE and is the answer not in the manual?

LINET Multicare LE Specifications

General IconGeneral
BrandLINET
ModelMulticare LE
CategoryMedical Equipment
LanguageEnglish

Related product manuals