42
11.10 Accessories
Risk of injury due to incompatible accessories!
Use original accessories from the manufacturer only.
NOTE The manufacturer is not responsible for the use of unapproved accessories.
11.10.1 Lifting Pole
To ensure safe use of the lifting pole:
Never exceed the maximum load of 75 kg.
Never use the lifting pole for rehabilitation exercises.
To prevent the bed from tipping over, ensure that the lifting pole does not project out from the bed.
Replace plastic handle every 4 years.
To install the lifting pole:
Insert lifting pole in corresponding sleeve fitting on lifting pole adapter at head end.
Ensure that safety pin locks into place.
A plastic grab handle with an adjustable strap shall be attached to the lifting pole.
NOTE The lifting pole adapter is optional. It is necessary to specify this feature in the order.
NOTE The date of manufacture is marked on the grab handle. LINET
®
recommends that you replace the
plastic grab handle every four years.
11.10.2 Infusion Stands
Infusion stands can be fitted to the head and foot end of the bed by either fitting into the IV/Infusion sockets
mounted on the bed or using alternative accesory holder socket in the head end on the undercarriage of the bed.
Use exclusively infusion stands with 4 hooks for hanging IV bags or baskets for intravenous solutions.
Ensure the infusion stand individual hook 2kg maximum Safe Working Load is not exceeded.
Capacity per hook: 2 kg (4.41 lbs).
Ensure the infusion stand 20kg maximum Safe Working Load is not exceeded.
The total maximum loading of the IV/Infusion poles must not exceed 20 Kg (44.1 lbs).
Risk of injury due to use of incorrect accessories or because of incorrect use!
Infusion Stands must only be used for their intended use. Always read the instructions for use!
Only mount an infusion pump to the lower (wider) telescopic section of an infusion stand
above the head/foot end board.
Never mount an infusion pump to the upper (thiner) telescopic section of an infusion
stand.
Ensure the infusion pump will not collide with any movable parts of the bed (especially
backrest part) or with the patient. This must be verified after installation.
Do not over tighten the infusion pump clamps during fitment. Over tightening may
damage the infusion stand.
Infusion pump can be only used if the infusion stand is fitted in the accessory holder
socket in the head end on the undercarriage of the bed (see Fig. 17b).