Danger to life due to electric shock!
If a fault occurs, have the electric motor, power box or other
electrical parts repaired by qualified personnel only.
Do not open the protective covers of the electric motor or the power
box.
23
Troubleshooting
Error/Fault
Cause
Solution
Adjusting with position buttons not
possible
GO button was not pressed
Press the GO button.
Function disabled on supervisor
panel
Enable disabled function.
Drive motors have no power
Defective drive motors
Defective battery
Check the mains connection.
Notify the service department.
Plug inserted incorrectly
Insert the mains plug correctly.
Faulty power source.
Notify the service department.
Faulty control element
Notify the service department.
Faulty mattress platform height/tilt
adjustment
There is an object on the undercar-
riage cover
Remove the object.
Function disabled on supervisor
panel
Enable disabled function.
Drive motors have no power
Defective drive motors
Defective battery
Check the mains connection.
Notify the service department.
Plug inserted incorrectly
Insert the mains plug correctly.
Faulty power source
Notify the service department.
Faulty control element
Notify the service department.
Lowering backrest from the upright
position not possible
There is an object under the back-
rest or in the drive mechanism
Remove the object.
Locking handle is defective
Notify the service department.
Adjusting siderails not possible
The siderail lock is dirty
Clean the locking mechanism.
Locking handle is defective
Notify the service department.
Faulty brakes
Dirt blocking brakes mechanically
Clean the brake system.
The brake mechanism is defective
Notify the service department.
Mattress not inflating
SCU mains switch turned off
No power to bedframe
Mains plug inserted incorrectly
Faulty power source
Faulty SCU
Air leaking
CPR valve leaking or open
Air pipes blocked
Mattress partialy inflated
Mattress damaged or faulty
Turn switch on. Green mains power
switch will illuminate.
Check mains connection.
Insert mains plug correctly.
Notify the service department.
Notify the service department.
Check mattress pipe connection.
Check CPR valves closed.
Check piper are not trapped or.
kinked
Check mattress is unfolded and flat.
Notify service department.
Unable to change mode or pressure
Go button was not pressed
Press the GO button.
Fault symbol illuminated & audio
alarm
One time exception fault
Persistent reoccuring fault
Switch off SCU & switch back on,
reset to see if failt is self-cleared.
Notify service department, mute
audio alarm.
CPR symbol illuminated
In CPR mode
Press CLP or Max to cancel

Table of Contents

Other manuals for LINET Multicare LE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare LE and is the answer not in the manual?

LINET Multicare LE Specifications

General IconGeneral
BrandLINET
ModelMulticare LE
CategoryMedical Equipment
LanguageEnglish

Related product manuals