81
Electrical safety
30.5 Mechanical Specifications (Symbioso)
30.6 Electrical Specifications (Symbioso)
Supply voltage
Model 230 V
220 - 240 V
50/60 Hz
Nominal power
Model 230 V
max. 20 VA (when operating from mains supply)
Fuse
Model 230 V
2x T1AH anti-surge fuse
Electrical safety class
Class 1 with applied parts type B
In conformity with EN 60601-1
NOTE Upon request, Linet
®
can deliver hospital beds with electrical specifications that comply with
regional standards (custom voltage, different mains plugs).
Identification of applied parts (Type B)
mattress platform frame, covers and all movable parts
head and foot end
siderails
Mobilift handles
Handset
Dimensions
Mattress (inflated)
SCU
213 cm x 86 cm x 20 cm
36 cm x 22 cm x 10 cm
Weight
Mattress (inflated)
SCU
9.5 kg
3.5 kg
Inflation time after storage
CPR deflation time
15 min
max. 30 s (electric or manual)
Environmental conditions - Operation
+10 °C +40 °C
Temperature
Humidity
30 75%
Atmospheric Pressure
79 106 kPa
Environmental conditions - Storage and Transport
Temperature
-40 °C — +70 °C
Humidity
10 100% (non-condensing)
Atmospheric Pressure
79 106 kPa
Max. mattress load
250 kg/550 lbs
Remains inflated in Transport Mode for
min. 12 hours
(when starting from Max Inflate Mode)
Noise level
NC30 (suitable for use in quiet domestic environment)
max. 45 dBa (normal operation without alarm)

Table of Contents

Other manuals for LINET Multicare LE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare LE and is the answer not in the manual?

LINET Multicare LE Specifications

General IconGeneral
BrandLINET
ModelMulticare LE
CategoryMedical Equipment
LanguageEnglish

Related product manuals