29
Status “Faulty battery”
The battery is regarded as faulty if at least one of the following conditions applies:
Battery charging constantly
Low voltage on battery
Low charging current of battery
This status is indicated by the battery status indicator being constantly lit.
These statuses are summarised to Linis and written to Blackbox.
To cancel this status:
Press STOP button.
Status “Discharged battery”
The battery is regarded as discharged if the following condition is met:
Defined decrease of voltage depending on discharging current
This status is indicated by the battery status indicator flashing quickly.
The electric CPR position is the only possible position.
This status will be cancelled automatically when the bed switches to sleep mode.
To cancel this status:
Press STOP button.
10.4 Removing the Bed from Service
Remove the bed from service as follows:
Disconnect the bed from the mains.
Disconnect the ground wire.
Deactivate the battery.
Remove accessories.
To prevent damage during storage:
Pack or cover the bed and accessories.
Ensure that storage conditions are the same as the operating conditions.
10.4.1 Deactivating the Battery
To avoid damaging the bed and the environment during storage:
Deactivate the battery on the supervisor panel.
Disconnect the bed from the mains.
Disconnect the ground wire.
Activate the keypad by pressing the GO button on the supervisor.
Press the Thigh Rest Up + Thigh Rest Down + Trendelenburg Position buttons at the same time and hold
them for three seconds.
The battery is deactivated.