Additional Functions of Symbioso
19.1 Transport Mode/Power Failure
This mode is activated automatically if no mains power is available to the SCU. CLP Mode will not be
operational.
The mattress maintains sufficient air pressure to support the patient for approximately 12 hours. The foam base
makes sure that patient does not lie directly on the mattress platform once the mattress deflates.
Possible reasons for SCU not being supplied with mains power:
Auxiliary power cord unplugged to move bed with patient
Power failure
Provide Multicare with Symbioso with mains power as soon as possible.
19.2 CPR (Symbioso)
Fig. 37 CPR Strip
Symbioso 100 is equipped with CPR pull tags on both sides next to the manual backrest release.
Pull CPR pull tags on patient’s left- or right-hand
side.
The mattress will deflate.
CPR Mode is activated.
Before re-inflating the mattress:
Unzip the cover, but do not remove the cover.
Reconnect the CPR valve sealing caps, mounted internally on both sides of the mattress at the head
end.
Close the cover and ensure that the red CPR pull tags are hanging outside the cover through the slots
in the cover.
NOTE In the pictures, the mattress is shown without the cover for clarity. Removing the cover is not
neccessary
to reconnect the CPR valves and replace the CPR pull tags.
Risk of injury due to lack of support!
The patient is not receiving CLP prevention while the power is off and
should be returned as soon as possible.