Max Inflate Mode (MAX)
Max Inflate Mode guarantees a firm surface as required for nursing procedures. This mode interrupts CLP
(Cons- tant Low Pressure).
Max Inflate Mode is required for:
transferring patients
complex nursing procedures
transporting the bed
To activate/deactivate the mode:
Press button 15.
The mattress icon will flash during the inflating process, and remain ... as soon as maximum inflation is obtained.
After 30 minutes, the SCU will automatically switch back to CLP mode. It is possible to re-select Max Inflate Mode
once. After that, at least 30 minutes of CLP mode are required before Max Inflate Mode is enabled again.
15
Constant Low Pressure Mode
CLP Mode keeps the mattress pressure at the level selected. The pressure is checked every 30 seconds, and
adjusted if necessary.
To activate/deactivate the mode:
Press button 16.
15.1 Fowler Boost Function
The Fowler Boost function linearly increases pressure in the seat section according to position of back rest.
It is possible to disable this function for lighter patients.
To enable/disable the function:
Press button 13.
16
MCM Mode
The MCM mode is the default mode for the Symbioso 200 mattresses, as this is the most effective mode for the
patient in view of the clinical effect. The MCM function blows through the parts under the patient and removes
moisture as one of the factors contributing to the development of bed sores. You can switch to the MCM mode by
pressing icon 11 (Fig. 33).