18.9 Electrical specification
Specification
Value
Battery Voltage
36 V DC, Capacity: 12 Ah
Maximum Power Input
300 W
Fuse
Accumulator fuse
pipe fuse T 3.15 A
MDP 030 (30 A)
18.10 I-Drive Power Maintenance
Periodical maintenance of the i-Drive Power must be done by qualified service technician or authorized service
organization at least once a year.
Service technician must check the following:
battery status and eventual replacement of batteries (after maximum of three years of duty)
gas spring replace if necessary (after maximum of three years of duty)
i-Drive Power wheel replace if necessary
lifting mechanism grease if necessary
cables, control elements replace if necessary
i-Drive Power function
NOTE To continue maintenance please see chapter Maintenance.