28
Damage to the bed due to incorrect battery replacement!
Have the battery replaced only by qualified personnel.
only use batteries approved by the manufacturer.
Battery capacity is sufficient (charging completed)
Short flashing (shortly lit,
longer not lit) (circa 1.8 sec.)
Battery is charging - continue charging until the LED is extinguished. In emer-
gency cases, the battery can be used as a backup power source for a short
period. If LED is still flashing after 12 hours of charging or stops flashing, but
you can not position with bed, battery is defective or broken. Contact manufac-
turer.
Long flashing (longer lit,
shortly not lit) (circa 0.2 sec.)
Low battery voltage - battery can not be used as a backup power supply even
for a short period; battery is completely discharged or defective (if this type of
signalisation persists, it is necessary to replace the battery - service action)
Lit continuously for several
hours (circa 10 hours), when
bed is connected to the mains.
The LED indicates the battery’s charge status:
Battery absence or failure condition (battery is connected incorrectly, line
between the power supply and battery is broken or battery fuses are faulty);
contact service department of the manufacturer in case of such signalisation.
10.3.1 Replacing the battery
Have batteries replaced only by a qualified service organisation.
For more detailed information on how to replace the batteries, request service manual from manufacturer.
Risk of reducing battery durability due to incorrect use!
Use bed on battery only in crisis situations (e.g.: power blackout, patient
compli- cations during transport, etc.)
After reconnecting bed to the mains charge battery to full capacity (see chart
Bat- tery charge status)
Material damage due to overheating!
If the battery is faulty, degassing may occur. In rare cases this might cause
deformations of the battery case, control panel housing or cable.
Stop using the bed immediately (see Removing the Bed from Service).
Inform the manufacturer’s service department.