24.3.2 Completeness
Perform a visual check (with delivery note if necessary).
Have any missing parts replaced.
24.3.3 Wear
Check all bolts and tighten if necessary.
Check all locking mechanisms.
Check the bed for wear, scratches or rub marks.
Eliminate the cause if necessary.
Have any defective parts replaced.
24.3.4 Functioning
Check that all bed adjustments reach the maximum position.
If necessary, clean, lubricate or replace any worn spots and parts.
24.3.5 Electric Control
Plug Connections:
Replace O-rings on connectors.
Check plug connections for dirt and defects.
Clean or replace if necessary.
Check that the plug connectors are properly seated.
Motors:
Check motor movement (adjust bed positions).
Check for incorrect and interrupted movements.
Have defective motors replaced if necessary.
Check cables for signs of wear and entanglement.
Install a new cable or have it replaced if necessary.
Battery:
Check that the battery is working properly (disconnect the bed from the mains).
Have the battery replaced if necessary.
Fuses:
Have fuses changed only by qualified and trained service technicians authorized by the manufacturer.
Use the following fuse types only:
T2A (for 230 V input)
T4A (for 100 – 127 V input)