26
10. Plug-In Power Supply, Specifications
Type: Model FJ-SW2401000N
Input voltage: 100-240V AC, 50/60 Hz, max 0.5 A.
Output: 24V DC, max 1.0 A
Ambient temperature: 0-40°C (+32°F to +102°F)
Dimensions (LxWxD): 82.4 x 44.5 x 36.2 mm / 3.2” x 1.8” x 1.4” + input plug
It is also possible to order a hardwired power supply option when and were it is needed.
11. Environmental conditions for the system
a) For indoor use.
b) Altitude operating range to 5500m (18000 ft).
c) Ambient temperature 0°C to +40°C.
d) Maximum relative humidity 95% (non condensing).
e) Mains supply voltage fluctuations up to ±10% of the nominal voltage.
f) Transient overvoltages up to the levels of overvoltage category II.
NOTE: These levels of transient overvoltage are typical for equipment
supplied from the building wiring.
g) Pollution degree 2.
12. Service and maintenance
1. Should be performed only by authorized professional service agents who are familiar with the O2
Safety System and all pertinent safety and service procedures. Contact your representative for
the name of the authorized service agent(s) in your area.
2. Since this is a safety product we recommend that a function check be performed on the O2
Safety System by a qualified professional service agent at least once every year.
3. The O2 Safety System has no user serviceable parts. All service work should be performed by an
authorized professional agent.
4. NOTE: Any attempt to service the equipment by unauthorized persons or to perform
unauthorized modifications will void the warranty.
5. The O2 sensor and central unit housing must NEVER be opened by unauthorized personnel.
6. Cleaning is done by use of water on a moistened cloth.