Channel Calibration
- 20 -
MXD70 Auxiliary mA Input
Setup and Operating Guide
Channel Calibration
Mode
Selecting off-line causes any setpoints associated
with this channel to de-energise. Also causes any
current outputs associated to hold their current
value, useful for when commissioning or calibrating
the instrument.
When the unit is placed in an off-line state “off-line”
will appear in the channel messages section on the
front screen.
If a “Cannot Edit Digital Input Has Control” message
appears, then an associated digital input is currently
controlling the on-line / off-line state of the channel.
/ – Select Option
EXIT
– Cancel
– Save Selection
Auxiliary mA Input Calibration
Enter the Auxiliary mA Input Calibration routine.
Allows the user to calibrate the mA input to a known
current source.
/ – Select Option
EXIT
– Return to Select Calibration Channel
– Enter pH Auto Calibration
0 mA Input Calibration
Set the known current input to 0mA and press enter
to initiate a calibration.
Only available when the input mode is set to 0 –
20mA in the channel setup menu.
PREV
– Go to Previous Calibration Point
SKIP
– Skip to Next Calibration Point
EXIT
– Exit Calibration Without Saving
– Initiate Calibration