BE6043-16 Stand Alone Multipoint System LumaSense Technologies A/S
Page 12 of 25
4) Select the Normalization temperature
The default Normalization temperature is 20˚C. You can select between
0˚C, 20˚C and 25˚C.
4.4 Set-up Alarm Limits
4.4.1. Alarm limits for a monitor without the Analog/Relay Module
UA1373 or UA1374
The alarm limits of a gas are those concentrations of gas which
should activate any alarm relay connected to the Monitor via its back
panel “Alarm Relay” socket. There are two different alarm relays,
which function slightly differently. Details of these relays and the way
in which they function is described in Chapter 4.4.2.The alarm level
parameters are found in the Gas Monitor…Configuration…Filters
branch of the Set-Up Tree (see Figure 3).
Warning! If the UA number for any filter is changed here, then all
the calibration data for that filter will be lost and the filter will need
to be recalibrated.
Follow the Set-up tree in Figure 3 until the following is displayed:
1. Use the select push-buttons to ignore this filter or to select this
filter if alarms are to be set.
2. Press 5 times until the following text is displayed.
3. Use the direction push-buttons to set the alarm concentration.
If you do not wish to activate the alarm function for a particular
gas, select the DEFAULT. This will “blank out” the alarm level
value for the gas and deactivate the alarm for that gas.