Operation Manual easyTymp
8111254 Rev. 9
15
17/01/2018
3.3.2 Cleaning the Case and Cables
Use caution while cleaning.
Use a damp cloth to clean the plastic parts of the easyTymp.
If disinfection is required, use a disinfectant wipe rather than
a spray product. Make sure that excess liquid from the wipe
does not seep into any sensitive areas such as connectors
and seams where plastic pieces connect such as the edges
around the touch screen.
Follow the instructions on the disinfection product.
3.3.3 Cleaning the probe tip
In order to secure correct impedance measurements it is important to make sure that the
probe system is kept clean at all times. Therefore please clean the probe on a periodic
basis. It is indispensable to remove cerumen from the probe tip’s small Acoustic and air
pressure channels. Therefore please follow the illustrated instructions below.
Figure 2
Never clean the probe tip while the tip is still
attached to the probe (Figure 2).
Figure 3
1. Unscrew the probe cap by turning it in a
counter clockwise direction (Figure 3).
Figure 4
2. Take the plastic probe tip out of the probe
(Figure 4).
CAUTION

Other manuals for Maico easyTymp

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maico easyTymp and is the answer not in the manual?

Maico easyTymp Specifications

General IconGeneral
BrandMaico
ModeleasyTymp
CategoryMedical Equipment
LanguageEnglish

Related product manuals