VIBRATION TRANSMITTER-VT7S12E
Ref No: mVTD/om/201
Issue No: 02
User’s Manual - 15 -
Channel skip/
Unskip selection.
RGP.4 / rGP.2
0:Relay Group-4
1:Relay Group-2
2:Relay Group-1
9 6 0 0 / 1 9. 2 K
0:(9 6 0 0 ) – 9600 bps
1:(1 9 . 2 K ) –19.2 Kbps
Parity/Stop bit
selection
PnS1 / PnS2 / PoS1 / PES1
0:(P n S 1 )-parity none-stop bit-1
1:(P n S 2 )-parity none-stop bit-2
2:(P o S 1 )-parity odd-stop bit-1
3:(P e S 1 )-parity even-stop bit-1
Timeout for
display back to
Run Mode
Retransmission
Open sensor
Retransmission
Output Type
0-20/4-20/ 0-5v/1-5v/
0-10v
0:(0 - 20) – 0-20mA
1:(4 - 20) – 4-20mA
2:(0 - 5 ) – 0 – 5volt
3:(1 - 5 ) – 1 – 5volt
4:(0 - 10) - 0 -10volt
Dir / rev
1:(dir)
0: (rev)
Retransmission
Channel Value
Max / Min
1:(Max)
0: (Min)
Password
Set password to
lock selected
level
Table 11 Level – 3 Parameter Description
Note:-
1) None means no AO(Analog Output)(Retransmission Output) is assigned to particular channel.
2) AO O/p number = 0 represents “None”. AO O/p Number = 1 and 2 represents “Retransmission No. 1 and Retransmission No.
2.
3) In “Retransmission mapping” parameter, maximum 2 numbers of Analog Outputs can be assigned to each channel. I.e. each
channel can have maximum two numbers of analog outputs simultaneously. See the example below.
Example 1: If Retransmission mapping for channel 1 = 0102 H, it means Retransmission output number 1 and output number
2 are assigned to channel 1. And output of channel 1 can be taken on analog output (retransmission output) number 1 and
number 2.
Example 2: If Retransmission mapping for channel 2 = 0001 H, it means Retransmission output number 1 is assigned to
channel 2. And output of channel 2 can be taken on analog output (retransmission output) number 1.