EasyManua.ls Logo

Masimo Rad-67 Pulse CO-Oximeter - User Manual

Masimo Rad-67 Pulse CO-Oximeter
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Operator's Manual
Rad-67™ Pulse CO-Oximeter
®
Spot-Check

Table of Contents

Questions and Answers

  • K
    Kristin CooperAug 20, 2025
    What to do if parameters are dimly lit on Masimo Medical Equipment?
    • M
      micheal96Aug 20, 2025
      If the parameters on your Masimo Medical Equipment are dimly lit, it could be due to low signal quality, so assess the patient. You can also try the following: * Verify the sensor type and size and re-apply the sensor. * Check if blood flow to the sensor site is restricted. * Check the placement of the sensor. Re-apply sensor or move to a different site. * Replace the sensor. * Minimize or eliminate motion at the measurement site. * Set to MAX Sensitivity.
  • S
    sarahmillerAug 21, 2025
    What to do for Low SIQ message displayed on Masimo Rad-67 Pulse CO-Oximeter Medical Equipment?
    • D
      Dr. Jason EstesAug 22, 2025
      If you see a Low SIQ (signal quality) message on your Masimo Medical Equipment, try the following: * Verify the sensor type and size and re-apply the sensor. * Check if blood flow to the sensor site is restricted. * Check the placement of the sensor. Re-apply the sensor or move to a different site. * Replace the sensor/cable. * Minimize or eliminate motion at the measurement site. * Set to Maximum Sensitivity.
  • A
    adamgarciaAug 25, 2025
    Why touchscreen/buttons do not respond when pressed on Masimo Rad-67 Pulse CO-Oximeter Medical Equipment?
    • Z
      Zachary WhiteAug 25, 2025
      If your Masimo device's touchscreen or buttons aren't responding, it could be due to EMI (Electro Magnetic Interference). Try turning the device off and on. Also, relocate the device away from other devices that may cause electromagnetic interference.
  • S
    Sherri SawyerAug 26, 2025
    Why is it difficult obtaining a reading on my Masimo Medical Equipment?
    • R
      rebeccathomasAug 26, 2025
      If you're finding it hard to get a reading on your Masimo Medical Equipment, try these steps: * Give the parameter reading some time to stabilize. * Make sure the sensor type and size are correct, and re-apply the sensor. * Ensure blood flow to the sensor site isn't restricted. * Check the sensor's placement; re-apply it or try a different site. * Replace the sensor, if necessary. * Confirm that the device and sensor are set up with the correct parameter. * Use the right sensor and size for the patient. * Shield the sensor from strong or flashing light. * Reduce or stop movement at the measurement site. * Check and adjust the 50 or 60Hz menu setting.
  • T
    Thomas Aguilar PhDAug 27, 2025
    What to do if SpHb reading is dim and 'Low SpHb SIQ' message displays on Masimo?
    • J
      Janet BurchAug 27, 2025
      If the SpHb reading is dim and you see a 'Low SpHb SIQ' message on your Masimo device, try the following: * Verify and re-apply the sensor. * Check if blood flow to the site is restricted. * Check the sensor's placement; re-apply it or move it to a different site. * Replace the sensor. * Minimize or eliminate motion at the monitoring site. * Submit a blood sample for a laboratory CO-Oximetry test for comparison. * Check patient conditions indicated to affect SpHb accuracy.
  • J
    Jonathan SolisAug 29, 2025
    Why are parameter readings displayed as dashes '- -' on my Masimo Rad-67 Pulse CO-Oximeter?
    • J
      jacksonlynnAug 29, 2025
      If your Masimo Medical Equipment displays parameter readings as dashes '- -', consider these steps: * Allow time for the parameter reading to stabilize. * Verify the sensor type and size, then re-apply the sensor. * Check if blood flow to the sensor site is restricted. * Check the sensor's placement; re-apply the sensor or move to a different site. * Replace the sensor. * Verify proper sensor and sensor size for the patient.
  • R
    Rachel CookAug 31, 2025
    What to do if device screen is blank on Masimo Medical Equipment?
    • K
      Kristin StevensAug 31, 2025
      If your Masimo Medical Equipment's screen is blank, try the following: * Touch the screen to turn the display on. * Turn the device off and on. * Adjust the brightness setting. * Check the AC power connection and recharge, as the battery may be depleted.
  • S
    Stephen MasonSep 2, 2025
    Why does speaker not work on Masimo Rad-67 Pulse CO-Oximeter Medical Equipment?
    • J
      Jason JenningsSep 2, 2025
      If the speaker on your Masimo device isn't working, it might be due to incorrect device audible settings. Try turning the device off and on. Also, ensure that alerts and sounds haven't been silenced and check the alert and sound volume settings. Finally, make sure the device speaker isn't muffled.
  • B
    Brittany OconnorSep 5, 2025
    How to fix a dim device screen on Masimo Medical Equipment?
    • D
      Diane SmithSep 5, 2025
      If your Masimo Medical Equipment's screen is dim, try touching the screen to return the display to full brightness. You can also adjust the brightness setting.
  • D
    david72Sep 7, 2025
    What to do if device does not detect that patient cable is connected to Masimo Medical Equipment?
    • G
      gilltraciSep 7, 2025
      If your Masimo device does not detect that the patient cable is connected, try removing and reconnecting the cable. Ensure the connector is fully connected to the device. If the issue persists, replace the cable.

Summary

Safety Information, Warnings, and Cautions

General Safety Warnings

Warnings about device use, modification, placement, unauthorized devices, and flammable environments.

Performance and Compliance Warnings

Warnings on performance, medical decisions, SpHb use, defibrillation, compliance, and modifications.

Cleaning and Service Cautions

Warnings about cleaning, maintenance, and avoiding electric shock.

Chapter 3: Setting Up

Initial Setup Procedures

Unpacking, inspection, preparation, guidelines, and initial battery charging for Rad-67.

Powering the Rad-67 ON and OFF

Procedures for turning the Rad-67 device on and off.

Chapter 4: Operation

Using the Touchscreen Interface

How to use the touchscreen, home button, gestures, and controls for device navigation.

Main Screen and Status Bar Overview

Explanation of the main screen layout, status bar details, and power indicators.

Parameter and Device Settings

Configuring SpO2, PR, SpHb, Pi settings, and general device settings like Localization and Wi-Fi.

Chapter 5: Spot-Check

Chapter 6: Alerts and Messages

Chapter 7: Troubleshooting

Chapter 9: Service and Maintenance

Masimo Rad-67 Pulse CO-Oximeter Specifications

General IconGeneral
BrandMasimo
ModelRad-67 Pulse CO-Oximeter
CategoryMedical Equipment
LanguageEnglish

Related product manuals