Operator's Manual
SedLine
®
Sedation Monitor

Table of Contents

Other manuals for Masimo sedline

Questions and Answers

  • A
    Aaron BlairAug 20, 2025
    What to do if the Masimo Monitor displays 'Sensor Off Patient'?
    • D
      Doris MyersAug 20, 2025
      If the Masimo Monitor displays 'Sensor Off Patient', ensure that the CB and CT are properly connected and the sensor is correctly inserted into the patient cable connector. If the issue continues, the sensor, patient cable, or module might be defective and require replacement.
  • M
    Matthew LewisAug 22, 2025
    What to do if Masimo Monitor says 'Replace Sensor'?
    • D
      David ChambersAug 22, 2025
      If the Masimo Monitor prompts 'Replace Sensor', first, check that the sensor is correctly inserted into the patient cable connector. If the sensor has been in use for more than 24 hours, replace it. Also, verify that the sensor's expiration date has not passed.
  • L
    lfernandezAug 25, 2025
    What to do if the Masimo Monitor displays 'No Sensor Connected'?
    • M
      Melanie SuttonAug 25, 2025
      If the Masimo Monitor indicates 'No Sensor Connected', verify that the sensor is properly inserted into the patient cable connector. If the problem persists, the sensor, patient cable, or module may be defective and need to be replaced.
  • I
    icrosbyAug 26, 2025
    How to fix 'High Impedance' error on Masimo Monitor?
    • G
      Gina MirandaAug 27, 2025
      To resolve a 'High Impedance' error on the Masimo Monitor, first confirm that all electrodes of the sensor are properly connected. If the problem persists, the sensor may need to be replaced.
  • B
    Brandon BartlettAug 27, 2025
    How to resolve 'Gel Bridging Detected' on Masimo Monitor?
    • C
      Corey WilliamsAug 28, 2025
      To resolve a 'Gel Bridging Detected' message on the Masimo Monitor: 1. Clean any gel that has leaked outside of the electrodes on the patient's forehead. 2. Confirm that all electrodes of the sensor are properly connected. 3. If the issue persists, the sensor may need replacement.
  • N
    Nathaniel MartinAug 29, 2025
    How to fix 'SedLine is Disconnected' error on Masimo Monitor?
    • K
      Kathleen FinleyAug 30, 2025
      To resolve a 'SedLine is Disconnected' error on the Masimo Monitor, reconnect the module.
  • D
    David ClarkAug 31, 2025
    What to do if electrode icons are green or yellow and ARTF display shows artifacts on Masimo sedline?
    • K
      Kathleen WrightAug 31, 2025
      If the electrode icons on the display of your Masimo Medical Equipment are green or yellow and the ARTF display shows artifacts, first determine if the artifact is indeed the problem by checking the ARTF display. If it is, ask the patient to relax and close their eyes. If the issue continues, consult the Line Power and High Frequency Artifacts section.
  • K
    Kevin JohnsonSep 2, 2025
    What to do if EEG display does not display waveforms for electrodes on Masimo sedline?
    • R
      Rebecca VargasSep 2, 2025
      If the EEG display on your Masimo Medical Equipment doesn't show waveforms for electrodes, it means the electrode didn't pass the initial impedance check. Here's what you can do: 1. Push and wiggle the white lining around the electrodes to improve contact, but avoid pressing directly on the electrode to prevent gel leakage. 2. Remove the sensor from the patient. 3. Wipe the patient's forehead with alcohol and let it dry. 4. Apply a new sensor. For detailed instructions, refer to the Directions for Use (DFU) for the sensor.
  • M
    Manuel WilliamsSep 4, 2025
    What to do if Sensor Off Patient on Masimo sedline Medical Equipment?
    • M
      Michael DavisSep 4, 2025
      To address a 'Sensor Off Patient' message on your Masimo Medical Equipment, follow these steps: 1. Ensure that the CB and CT are properly connected. 2. Verify that the sensor is correctly inserted into the patient cable connector. 3. If the issue persists, the sensor, patient cable, or module may be defective and require replacement.
  • T
    timothy70Sep 6, 2025
    What to do if High Impedance on Masimo sedline Medical Equipment?
    • M
      Mary EllisSep 6, 2025
      To resolve a 'High Impedance' alert on your Masimo Medical Equipment, ensure that all electrodes of the sensor are properly connected. If the problem persists, the sensor may need to be replaced.

Summary

Product Information

Indications for Use

Specifies the intended clinical environments for the SedLine Monitor's application.

Contraindications

States the age restriction for using the SedLine Sedation Monitor.

Warnings and Cautions

Safety Information

Chapter 2: System Descriptions

SedLine Components

Details the SedLine Module, Patient Cable, and Sensor(s).

Chapter 3: Setting Up the System

Preparation for Use

Essential checks and confirmations required before operating the SedLine monitor.

Connecting Components

Instructions on connecting the module to the patient cable and Root.

Chapter 4: Operation

Chapter 5: Menu Options

Parameter Settings

Access and configure settings for various SedLine monitored parameters.

Additional Settings

Configure EEG and DSA settings, and SedLine Sensor information.

Chapter 6: EEG Download

Enable EEG Data Collection

Activates the recording and storage of EEG waveforms to a USB stick.

Chapter 7: Alarms and Messages

Chapter 8: Troubleshooting

Trend and Artifact Issues

Resolving issues with unavailable trends, impedance, and artifacts.

Interference Issues

Solutions for line power and high frequency interference artifacts.

Chapter 10: Service and Maintenance

Masimo sedline Specifications

General IconGeneral
Channels4 EEG channels
ParametersEMG
ConnectivityEthernet, USB
CompatibilityMasimo Root
Applicationsedation monitoring
Power SourceBattery