www.masimo.com 63 Masimo
Chapter 10: Service and Maintenance
Cleaning Procedures
Cleaning of SedLine should be performed at regular intervals in accordance with hospital, as
well as local and governmental regulations.
See the Warnings and Cautions section in this manual.
SedLine is a reusable instrument. The instrument is supplied non-sterile.
To clean the module
• The outer surface of the module can be cleaned with a soft cloth dampened with a
mild detergent and warm water solution.
• Do not allow liquids to enter the interior of the module.
• The outer surface of the module can also be wiped down using any of the following
solvents:
• Cidex Plus (3.4% glutaraldehyde)
• 10% bleach solution
• 70% isopropyl alcohol solution
To clean the patient cable
• Moisten a lint-free towel with a mild soapy solution or mild disinfectant. Do not
use abrasive cleaners
• Wipe down surfaces of the patient cable with the lint-free towel
• Dry completely after cleaning
General Maintenance for Module
Safety tests and internal adjustments should be done by qualified personnel only. Safety
checks should be performed at regular intervals or in accordance with hospital, as well as
local and governmental regulations.
The following is a checklist for the general maintenance of SedLine:
• Visually inspect equipment for functional or structural damage, including poor
seals, cracks, damaged springs, etc.
• Visually inspect cables, connectors, and connector pins for signs of damage or
wear.
• Visually inspect product identification labels to ensure they are clear and legible.