ATMe, Haze Generator
User Guide, Rev: C/e, March 2023 23 / 33
Symptoms Probable Causes and Suggested Actions
« STATE = FAIL»
VERIFY THE ERROR MESSAGES
ERROR = P. LOW
Open the gas bottle
With CO
2
bottle, verify the regulator is not
frozen
Verify the set pressure on the regulator (60
psi, 4.1 bar or 410 kPa)
Verify the gas line
Verify the reading of the pressure transducer
«STATUSPRESSURE»
Verify the regulator is not frozen
ERROR = P. HIGH
Verify the reading of the pressure transducer
«STATUSPRESSURE»
Verify you are not using a Siphon CO
2
bottle
You may have a solenoid valve malfunction.
Restart the generator.
Verify you are not using liquid CO
2
Please contact MDG
Service if this problem
persists.
ERROR = HEATER
This is a heating timeout, due to a cartridge
heater problem.
Restart the generator and check the heating
status «STATUSSTATE
= xx% HEAT». If
the heating value is not progressing, there
may be various issues.
Check the fuses on the heater board
Check the voltage
Please contact MDG Service if this problem
persists.
∕…