EN
11
05
Û
CLEANING AND STERILIZATION
PREPARATION
- Disconnect the foot pedal from the device
body (see Chapter 03)
CAUTION: Do not sterilize the foot
pedal in autoclave, it would stop working.
CAUTION: Do not clean the foot pedal
under running tap water.
CAUTION: Do not immerge the foot
pedal in liquids and/or solutions of various
nature.
MATERIALS NECESSARY
- Clean, soft cloths with low fiber-release
- Detergent solution (pH 6-9) and, if need be,
non-aggressive disinfectant solution with
neutral pH (pH7)
CAUTION: We recommend that you
use water-based disinfectant solutions
with neutral pH (pH7). Alcohol-based
disinfectant solutions and hydrogen
peroxide are contra-indicated, because
they can fade the color and/or damage the
plastic materials. This also holds true for
chemical products such as acetone and
alcohol.
CLEANING AND DISINFECTION METHOD - Foot pedal
Clean the surface of the foot pedal with
a clean, soft cloth with low fiber-release,
dampened with a detergent solution
(pH 6-9) and, if need be, disinfect with a
non-aggressive disinfectant solution with
neutral pH (pH7), following the instructions
provided by the manufacturer of the
solution;
Dry the foot pedal with a clean,
non-abrasive cloth with low fiber-release.
05.2
Û
CLEANING OF THE FOOT PEDAL
1
2
pH
6-9