EN
6
WARNING: Switch the device off.
Always turn the device off via the switch,
and disconnect the power supply cable
from the power outlet and from the
device body, before performing the
cleaning and sterilization interventions.
WARNING: Supplementary
grounding cable. If present, disconnect
the supplementary grounding cable
before you perform the cleaning and
sterilization interventions.
Before moving on to the cleaning procedures described in Chapter 05 - CLEANING AND
STERILIZATION, disconnect all the accessories and components of the PIEZOSURGERY® touch.
9
10
Once you have finished cleaning the irrigation circuit, move on to the cleaning and sterilization
of the individual parts (see Chapter 05 - CLEANING AND STERILIZATION).
To perform a cleaning cycle of every
handpiece, tube, or connector used during
the same treatment, repeat the operations
described above, from point 6 onward;
IMPORTANT After the cleaning cycle with
the enzymatic detergent solution, perform
a new cleaning cycle with distilled water to
remove detergent residues.
03
Û
DISASSEMBLY OF PARTS FOR CLEANING
AND STERILIZATION
1
off